Model 1007C Temperature Chamber Specifications


Temperature Range
-73°C to +175°C

Control Tolerance
±0.5°C (±0.2°C Typical) (Short-term variations measured at the control sensor after stabilization)

Uniformity
±1°C (Variations throughout the chamber after stabilization)
 
Cool Down Transition Time*
Start Temp End Temp
+23°C 0°C -40°C -55°C -65°C -73°C
+23°C ----- 4 min 18 min 25 min 33 min Ultimate
+50°C 5 min 11 min 27 min 36 min 45 min Ultimate
+85°C 13 min 20 min 37 min 47 min 55 min Ultimate
+150°C 32 min 39 min 58 min 65 min 76 min Ultimate
Heat Up Transition Time*
Start Temp End Temp
+23°C +50°C +85°C +125°C +150°C +175°C
+23°C ----- 1.5 min 7 min 14 min 20 min 25 min
0°C 1.5 min 3.5 min 13 min 20 min 23 min 31 min
-40°C 6 min 11 min 17 min 24 min 30 min 35 min
-55°C 8 min 13 min 19 min 26 min 32 min 37 min
-65°C 10 min 14 min 21 min 28 min 34 min 39 min

Rate Of Change
To calculate rate of change for a particular condition, take the difference between the Start Temp and End Temp and divide by the Transition Time.
Cool Down Example: From +50°C to -40°C = 90°C / 27 min = 3.33°C/min.
Heat Up Example: From -40°C to +50°C = 90°C / 11 min = 8.18°C/min.


*Note: Transition times are measured after a 2 hour soak at the respective start temperature with an empty chamber, as indicated on the temperature controller, 23°C ambient. Measured with setpoint beyond the start and end temperatures. Does not include the effect of proportional band when approaching setpoint. Performance is reduced by 17% with 50 Hz input power.

 
Live Load Capacity
+23°C 0°C -40°C -55°C -65°C
1,000 Watts 800 Watts 500 Watts 400 Watts 300 Watts
 
Refrigeration and Heating System
High Stage Refrigerant R-404A (Dupont HP-62)
Low Stage Refrigerant R-508B (Dupont SUVA-95)
Compressors 1.5 HP x 1.5 HP Tecumseh hermetic compressors in a cascade configuration.
Condenser Air Cooled
Heat of Rejection 14,800 BTUH (maximum rated chamber load at maximum cooling rate from high temperature soak)
Heater Power 2,000 Watts
Air Flow 450 cfm
 
Instrumentation
Temperature Controller Watlow F4T Touch Screen Controller with RS-232, Ethernet interface, 4.3" color graphic touch screen. OR...
Watlow F4 Controller with RS-232 interface, LED readout of temperature, LCD display of other parameters.
Limit Controller Independent high and low temperature limits. Triggers an audible alarm and shuts down the chamber. Relay contacts provide a safety power interlock for test sample.
Chart Recorder (Optional) Honeywell DR4300 Series. One pen, 10" circular chart. Mounts in lower front door. Honeywell Chart Recorder
 
Input Power Requirements
Phase Current Draw Minimum Service
230 V ±10%, 60 Hz 1 PH 25 A 30 A
3 PH 19 A 25 A
208 V -5/+10%, 60 Hz 1 PH 28 A 35 A
3 PH 21 A 30 A

Input may be configured for single or three phase in the field by changing jumpers. Three phase load is semi-balanced. Call for 50 Hz operation. Performance is reduced by 17% with 50 Hz input power.

Customer power source must be hard-wired to the chamber by a qualified electrician. Power cord and plug is not included.

 
Physical Characteristics and Safety
Inside Dimensions 24" W x 21" H x 24" D (7 cubic feet)
609 mm W x 533 mm H x 609 mm D (198 liters)
Outside Dimensions* 33" W x 67.75" H x 54.25" D (nominal)
838 mm W x 1721 mm H x 1378 mm D
* Door latch adds 3" to width on right side (may be removed to permit move-in through a 36" doorway). Circulator motor and housing adds 6" to height.
Minimum Installed Clearance 18" from the left and right side
24" from the rear
Window Viewing Area 13.375" W x 9" H
Access Ports 4" Port on left and right side (two total)
Supplied with foam plugs
Weight Chamber Weight: 850 pounds
Shipping Weight: 1,011 pounds
Sound Level 62 dBA in cooling mode (A-weighted, measured 36" from the front or side surface, 63" from the floor, in a free-standing environment)

NOTE: Performance is typical and based on operation at 23°C (73°F) ambient and nominal input voltage. Designed for use in a normal conditioned laboratory. Operation at higher ambient temperatures may result in decreased cooling performance. Additional ports and shelves will also affect performance. Operation above 30°C (85°F) or below 16°C (60°F) ambient is not recommended.