Model 1007H Temperature/Humidity Chamber Specifications
Standard Range: 10% to 95%
(Limited by a 6°C dewpoint and maximum dry bulb of +85°C)
With optional GN2 Purge or Dry Air Purge: 5% to 95%
(Dry bulb range of +10°C to +85°C)
NOTE: Ability to reach RH extremes may be limited by the humidity sensor accuracy. Low Dew Point conditions can only be achieved when starting with a clean, dry chamber.
Cool Down Transition Time* (uncontrolled humidity mode) | ||||||
Start Temp | End Temp | |||||
+23°C | 0°C | -40°C | -55°C | -65°C | -73°C | |
+23°C | ----- | 4 min | 18 min | 25 min | 33 min | Ultimate |
+50°C | 5 min | 11 min | 27 min | 36 min | 45 min | Ultimate |
+85°C | 13 min | 20 min | 37 min | 47 min | 55 min | Ultimate |
+150°C | 32 min | 39 min | 58 min | 65 min | 76 min | Ultimate |
Heat Up Transition Time* (uncontrolled humidity mode) | ||||||
Start Temp | End Temp | |||||
+23°C | +50°C | +85°C | +125°C | +150°C | +175°C | |
+23°C | ----- | 1.5 min | 7 min | 14 min | 20 min | 25 min |
0°C | 1.5 min | 3.5 min | 13 min | 20 min | 23 min | 31 min |
-40°C | 6 min | 11 min | 17 min | 24 min | 30 min | 35 min |
-55°C | 8 min | 13 min | 19 min | 26 min | 32 min | 37 min |
-65°C | 10 min | 14 min | 21 min | 28 min | 34 min | 39 min |
Rate Of Change
To calculate rate of change for a particular condition, take the difference between the Start Temp and End Temp and divide by the Transition Time.
Cool Down Example: From +50°C to -40°C = 90°C / 27 min = 3.33°C/min.
Heat Up Example: From -40°C to +50°C = 90°C / 11 min = 8.18°C/min.
*Note: Transition times are measured after a 2 hour soak at the respective start temperature with an empty chamber, as indicated on the temperature controller, 23°C ambient. Measured with setpoint beyond the start and end temperatures. Does not include the effect of proportional band when approaching setpoint. Performance is reduced by 17% with 50 Hz input power.
Live Load Capacity (uncontrolled humidity mode) | |||||
+23°C | 0°C | -40°C | -55°C | -65°C | |
1,000 Watts | 800 Watts | 500 Watts | 400 Watts | 300 Watts |
Refrigeration and Heating System | ||
High Stage Refrigerant | R-404A (Dupont HP-62) | |
Low Stage Refrigerant | R-508B (Dupont SUVA-95) | |
Compressors | 1.5 HP x 1.5 HP Tecumseh hermetic compressors in a cascade configuration. | |
Condenser | Air Cooled | |
Heat of Rejection | 14,800 BTUH (maximum rated chamber load at maximum cooling rate from high temperature soak) | |
Air Heater Power | 2,000 Watts | |
Humidifier Heater Power | 1,500 Watts (rating at 240V) | |
Air Flow | 450 cfm |
Instrumentation | ||
Temperature Controller | Watlow F4T Touch Screen Controller with RS-232, Ethernet interface, 4.3" color graphic touch screen. OR... Watlow F4 Controller with RS-232 interface, LED readout of temperature, LCD display of other parameters. |
|
Limit Controller | Independent high and low temperature limits. Triggers an audible alarm and shuts down the chamber. Relay contacts provide a safety power interlock for test sample. | |
Chart Recorder | (Optional) Honeywell DR4300 Series. One pen, 10" circular chart. Mounts in lower front door. |
Input Power Requirements | |||
Phase | Current Draw | Minimum Service | |
230 V ±10%, 60 Hz | 1 PH | 25 A | 30 A |
3 PH | 19 A | 25 A | |
208 V -5/+10%, 60 Hz | 1 PH | 28 A | 35 A |
3 PH | 21 A | 30 A |
Input may be configured for single or three phase in the field by changing jumpers. Three phase load is semi-balanced. Call for other voltages or 50 Hz operation. Performance is reduced by 17% at 50 Hz.
Customer power source must be hard-wired to the chamber by a qualified electrician. Power cord and plug is not included.
Humidity Water Requirements | |
Supply and Drain | Must be provided with a water line and floor drain. Negligible consumption. |
Water Recirculation System (optional) | Provides a reliable supply of filtered water for the humidity system. Perfect for installations where a water line and drain are not available. |
Physical Characteristics and Safety | |
Inside Dimensions | 24" W x 21" H x 24" D (7 cubic feet) 609 mm W x 533 mm H x 609 mm D (198 liters) |
Outside Dimensions | 33" W x 68.75" H x 54" D (nominal) 838 mm W x 1746 mm H x 1372 mm D |
Door latch adds 3" to width on right side (may be removed to permit move-in through a 36" doorway). Circulator motor and housing adds 6" to height. | |
Minimum Installed Clearance | 18" from the left and right side 24" from the rear |
Window Viewing Area | 13.375" W x 9" H |
Access Ports | 4" Port on left and right side (two total) Supplied with foam plugs |
Weight | Chamber Weight: 890 pounds Shipping Weight: 1,058 pounds |
Sound Level | 62 dBA in cooling mode (A-weighted, measured 36" from the front or side surface, 63" from the floor, in a free-standing environment) |
NOTE: Performance is typical and based on operation at 23°C (73°F) ambient and nominal input voltage. Designed for use in a normal conditioned laboratory. Operation at higher ambient temperatures may result in decreased cooling performance. Additional ports and shelves will also affect performance. Operation above 30°C (85°F) or below 16°C (60°F) ambient is not recommended.