Model 1007S Temperature Chamber Specifications
Cool Down Transition Time (empty)* | ||||||
Start Temp | End Temp | |||||
+23°C | 0°C | -10°C | -20°C | -30°C | -35°C | |
+23°C | ---- | 1.5 min | 3.5 min | 7 min | 13 min | Ultimate |
+85°C | 5 min | 10 min | 14 min | 18 min | 26 min | Ultimate |
+150°C | 12 min | 19 min | 22 min | 26 min | 33 min | Ultimate |
Cool Down Transition Time (with 25 lb. aluminum load)* | ||||||
Start Temp | End Temp | |||||
+23°C | 0°C | -10°C | -20°C | -30°C | -35°C | |
+85°C | 10 | 18 min | 22 min | 29 min | 41 min | Ultimate |
+150°C | 21 | 33 min | 38 min | 43 min | 56 min | Ultimate |
Heat Up Transition Time (empty)* | ||||||
Start Temp | End Temp | |||||
+23°C | +50°C | +85°C | +125°C | +150°C | +175°C | |
+23°C | ----- | 1.5 min | 7 min | 14 min | 20 min | 25 min |
0°C | 1.5 min | 3.5 min | 13 min | 20 min | 23 min | 31 min |
-10°C | 2.2 min | 4.2 min | 14 min | 22 min | 25 min | 33 min |
-35°C | 5.5 min | 10 min | 16 min | 23 min | 29 min | 34 min |
Rate Of Change *Note: Transition times are measured after a 2-hour soak at the respective start temperature with an empty chamber, as indicated on the temperature controller, 23°C ambient. Measured with setpoint beyond the start and end temperatures. Does not include the effect of proportional band when approaching setpoint. Performance is reduced by 17% with 50 Hz input power. |
Live Load Capacity | |||||
+23°C | 0°C | -10°C | -20°C | -30°C | |
1,400 Watts | 1,000 Watts | 850 Watts | 625 Watts | 300 Watts |
Refrigeration and Heating System | ||
Refrigerant | R-404A (Dupont HP62) | |
Compressors | 1.5 HP Tecumseh hermetic | |
Condenser | Air Cooled | |
Heat of Rejection | 15,200 BTUH (maximum rated chamber load at maximum cooling rate from high temperature soak) | |
Heater Power | 2,000 Watts | |
Air Flow | 450 cfm |
Instrumentation | ||
Temperature Controller | Watlow F4T Touch Screen Controller with RS-232, Ethernet interface, 4.3" color graphic touch screen. OR... Watlow F4 Controller with RS-232 interface, LED readout of temperature, LCD display of other parameters. |
|
Limit Controller | Independent high and low temperature limits. Triggers an audible alarm and shuts down the chamber. Relay contacts provide a safety npower interlock for test sample. | |
Chart Recorder | (Optional) Honeywell DR4300 Series. One pen, 10" circular chart. Mounts in lower front door. |
Input Power Requirements | |
230 V ±10%, 60 Hz, 1 PH |
Max Current Draw 22 A; Recommended Service 30 A |
208 V -5/+10%, 60 Hz, 1 PH |
Max Current Draw 22 A; Recommended Service 30 A |
Input may be configured for 230 V or 208 V in the field by changing jumpers. Single phase input only. Call for other voltages or 50 Hz operation. Performance is reduced by 17% with 50 Hz input power. Customer power source must be hard-wired to the chamber by a qualified electrician. Power cord and plug are not included. |
|
Physical Characteristics and Safety | |
Inside Dimensions | 24" W x 21" H x 24" D (7 cubic feet) 609 mm W x 533 mm H x 609 mm D (198 liters) |
Outside Dimensions* | 33" W x 67.75" H x 54.25" D (nominal) 838 mm W x 1721 mm H x 1378 mm D |
Door latch adds 3" to width on right side (may be removed to permit move-in through a 36" doorway). Circulator motor and housing adds 6" to height. | |
Minimum Installed Clearance | 18" from the left and right side 24" from the rear |
Window Viewing Area | 13.375" W x 9" H |
Access Ports | 4" Port on left and right side (two total) Supplied with foam plugs |
Weight | Chamber Weight: 755 pounds Shipping Weight: 916 pounds |
Sound Level | 62 dBA in cooling mode (A-weighted, measured 36" from the front or side surface, 63" from the floor, in a free-standing environment) |
NOTE: Performance is typical and based on operation at 23°C (73°F) ambient and nominal input voltage. Designed for use in a normal conditioned laboratory. Operation at higher ambient temperatures may result in decreased cooling performance. Additional ports and shelves will also affect performance. Operation above 30°C (85°F) or below 16°C (60°F) ambient is not recommended.