- Description
- Specifications
- Documents
Keysight 66321B/120 Mobile Communication DC Source
Configurations Include
- (120) DC power supply 104-127VAC operation, 50/60Hz
Features and Specifications
- Single Output
- Range: 15 V, 3 A
- Input: 120 VAC
- Interface: GPIB
- Battery Emulation
66321B Overview
The Keysight 66321B is a DC source for mobile communications that delivers fast transient response, exceptional sourcing, output resistance programming, and precision measurement. It is specifically engineered to test the products of future digital wireless communications. The 66321B introduces a novel feature that emulates the internal resistance of a battery or battery pack. Plus, these high-performance models provide enhanced transient response and stable output operation with either short or long load leads.
For R&D engineers, the 66321B enables the characterization of the battery's current output and facilitates design trade-offs that influence current drain and battery life. This is achieved by offering data logging (up to 1000 hours), CCDF statistical analysis, and waveform capture when the 66321B is used in conjunction with the Keysight 14565B software.
66300 Series Family of Specialized DC Power Sources
The 66300 series power supplies are specifically engineered to assist in testing mobile wireless devices. They offer DC sourcing, current sinking, and measurement functionalities to tackle the distinct difficulties of emulating batteries and battery packs, as well as gauging the current consumed by your test device.
Series Features
- Programmable output response compensation to accommodate all types of wiring configurations
- Fast DC power source to replace and simulate the battery during testing
- Fast voltage transient response ensures maximum test-system throughput by minimizing device shutdowns
- Current sinking for testing and calibrating charger circuitry
- Uses a proprietary system to automatically detect open sense wire connections
- Extensive protection features (including broken sense lead detection)
- GPIB Interface, SCPI (Standard Commands for Programmable Instruments), VXI plug&play drivers
- Dynamic measurement system enables accurate current measurement from µA to A
- 66319B/D and 66321B/D coupled with 14565B Software gives a powerful analysis tool to optimize device designs for long battery life
For scenarios where you need to substitute your device's primary battery with power during testing, single output models are suggested. On the other hand, dual output models are advised when you need to replace your device's main battery with power and simulate the power of the battery charger. In this case, one output is used to connect in lieu of the main battery (which sinks current to mimic the charging of the main battery), and the other output is used to deliver current to the battery charger's input port. Models are as follows:
Model | Power (W) | Max (V) | Max (A) | No. of Outputs | No. of Ranges | Ripple & Noise | Programming accuracy | Transient response ( |
66309B/D | 45 | 15 | 3 (5 A peak) | 2 | 1 | 6 mVp-p | 0.05% + 10 mV | < 35 |
66311B | 45 | 15 | 3 (5 A peak) | 1 | 1 | 6 mVp-p | 0.05% + 10 mV | < 35 |
66319B/D | 45 | 15 | 3 (5 A peak) | 2 | 1 | 6 mVp-p | 0.05% + 10 mV | < 20 |
66321B/D | 45 | 15 | 3 (5 A peak) | 1 | 1 | 6 mVp-p | 0.05% + 10 mV | < 20 |
66332A | 100 | 20 | 5 | 1 | 1 | 3 mVp-p | 0.05% + 10 mV | < 100 |
Feature | 66311B | 66309B | 66309D | 66321B | 66321D | 66319B | 66319D | 66332A |
---|---|---|---|---|---|---|---|---|
Fast response power technology | Yes | Yes | Yes | Yes | Yes | Yes | Yes | |
Peak current (up to 5 A) | Yes | Yes | Yes | Yes | Yes | Yes | Yes | |
Dynamic measurement capability (talk, stand-by, off modes), Measurement ranges | 2 | 2 | 2 | 3 | 3 | 3 | 3 | 2 |
Battery emulation capability | Yes | Yes | Yes | Yes | ||||
Replace main battery and power adapter (requires dual DC outputs) | Yes | Yes | Yes | Yes | ||||
Output response compensation, Programmable modes | 2 | 2 | 2 | 4 | 4 | 4 | 4 | 2 |
Electronic load capability for calibrating charger circuit A/D’s | Yes | Yes | Yes | Yes | Yes | Yes | Yes | |
Remote sense | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes |
Auxiliary built-in DVM | Yes | Yes | Yes | |||||
Broken sense lead protection | Yes | Yes | Yes | Yes | Yes | Yes | Yes | |
Output disconnect relays (Opt. 521) | Yes | Yes | Yes | Yes | Yes |
Series Size 1/2 RU w x 2 RU h: RU refers to rack unit of a standard 19-inch EIA equipment rack. The width is either 1/2 or full. The height is in number of rack units which are 1.75 inches (44.4 mm) each. For example: a 3 RU h has a height of 5.25 inches (133.3 mm).