- Description
- Specifications
- Documents
Fluke 1586A/2DS Super-DAQ and DAQ-STAQ Multiplexer
Features
- Measure thermocouples, PRTs, thermistors, DC voltage, DC current, and resistance
- Best-in-class temperature measurement accuracy
- PRTs ±0.005 C (using external DAQ-STAQ Multiplexer)
- Thermocouples: ±0.5 C (using High-Capacity Module and internal CJC)
- Thermistors: ±0.002 C
- Input channels: Up to 40 isolated universal inputs
- Selectable scan speed: Up to 10 channels per second
- Four modes of operation: Scan, Monitor, Measure, Digital Multimeter (DMM)
- Real-time color trending: Chart up to four channels simultaneously
- Current reversal minimizes thermoelectric effects, improving PRT/RTD measurment accuracy
- Control Fluke Calibration temperature sources such as dry-wells or micro-baths for automated calibration routines
- Records up to 20 MB of data and setup files to internal non-volatile memory or to an external USB drive
- Transfer data to a PC using the USB drive or LAN connection and view data in Microsoft Excel
- Data security: Administrator and user profiles for protecting settings and ensuring test traceability
- Mx + B scaling and channel offset zero function
- Two independent, user-defined alarms for each channel indicate when an upper or lower range has been exceeded
- 6 1/2 digit display resolution for DC voltage, DC current, and resistance
1586A Super-DAQ Precision Temperature Scanner
The 1586A Super-DAQ is the most accurate and flexible temperature data acquisition system on the market. It scans and records temperature, DC voltage, DC current, and resistance for up to 40 input channels and scan speeds as fast as 10 channels per second. The Super-DAQ can be configured for use as a multi-channel data logger in the factory or as a precision reference thermometer for benchtop sensor calibration in the lab.
When configured with the DAQ-STAQ Multiplexer, the Super-DAQ has the accuracy of the best benchtop reference thermometer readouts for calibration of PRTs, RTDs, thermistors, or thermocouples. Lab efficiency can be increased when the Super-DAQ is connected to a Fluke Calibration drywell or bath and Automated Sensor Test routines are run. Furthermore, pairing the 1586A Super-DAQ with a high-capacity module (sold separately) adds the ability to measure temperature, resistance, dc voltage, and dc current for data acquisition applications.
The 1586 is ideal for a number of applications such as thermal mapping, temperature validation, process sensor calibration, and more. These applications are found in a various industries including pharmaceutical, biotechnology, food processing, aerospace, and automotive.
General Specs
- Maximum input: 50 V
- Offset voltage: < 2 µV
- 3-wire internal resistance mismatch: <50 megohm
- Basic CJC accuracy: 0.25 Celcius
DAQ-STAQ Multiplexer
Designed for high-accuracy measurements in secondary temperature calibration labs. Easily connect/disconnect thermocouples, PRTs, and thermistors.
General Specs
- Maximum input: 50 V
- Offset voltage: < 2 µV
- 3-Wire internal resistance mismatch: < 50 megohm
- Basic CJC accuracy: 0.25 Celsius
For a calibration lab where accuracy is of primary importance, the Super-DAQ is best configured with a DAQ-STAQ Multiplexer. The external DAQ-STAQ features mini-jack thermocouple terminals - each with its own reference junction sensor - and patented mini-DWF, gold-plated input terminals, which accept bare wire, spade lug, or mini-banana-plug terminations. Easily connect and disconnect PRTs, thermistors, and thermocouples for benchtop temperature calibration. It can be stacked on the 1586 to reduce footprint in busy labs. Flexibility to configure the Super-DAQ for factory or calibration lab use reduces your equipment needs and cost.
What's in the Box
- 1586A/2DS Super-DAQ
- (2) DAQ-STAQ Multiplexer