TestEquity
Chambers Home

What is the Retransmit Output Option for Environmental Chambers?

  • Provides scaled analog output.
  • NOT a computer interface.
  • Typically used to drive an analog recorder.
  • Example: 0 to 5 VDC = 0 to 100% Humidity (therefore 2.5 V = 50% RH).
  • Example: 0 to 5 VDC = -75°C to +175°C (therefore 2.5 V = 50°C).
  • Typical use is to monitor actual Temperature and Humidity (on a humidity chamber) or actual Temperature and Temperature Setpoint.
  • Terminals are on the back of the F4 and F4T Controller.

Retransmit Option 0004 on an F4 Controller

  • User-selectable scale: 0 to 10 VDC, 0 to 5 VDC, 1 to 5 VDC, 0 to 20 mA, 4 to 20 mA.
  • Provides two outputs.

Retransmit Option F4T-RETRANSMIT on an F4T Touch Screen Controller

  • Separate scale low and scale high levels, up to 24 VDC output or 100 mA.
  • Provides three outputs.
  • Configurable using Composer Software
  • When used in a humidity chamber, eliminates two of the available event outputs.