TestEquity
Information notification
Due to ongoing tariff updates, pricing is subject to change and may not reflect the latest surcharges. Learn more.
Chambers Home

What is the Retransmit Output Option for Environmental Chambers?

  • Provides scaled analog output.
  • NOT a computer interface.
  • Typically used to drive an analog recorder.
  • Example: 0 to 5 VDC = 0 to 100% Humidity (therefore 2.5 V = 50% RH).
  • Example: 0 to 5 VDC = -75°C to +175°C (therefore 2.5 V = 50°C).
  • Typical use is to monitor actual Temperature and Humidity (on a humidity chamber) or actual Temperature and Temperature Setpoint.
  • Terminals are on the back of the F4 and F4T Controller.

Retransmit Option 0004 on an F4 Controller

  • User-selectable scale: 0 to 10 VDC, 0 to 5 VDC, 1 to 5 VDC, 0 to 20 mA, 4 to 20 mA.
  • Provides two outputs.

Retransmit Option F4T-RETRANSMIT on an F4T Touch Screen Controller

  • Separate scale low and scale high levels, up to 24 VDC output or 100 mA.
  • Provides three outputs.
  • Configurable using Composer Software
  • When used in a humidity chamber, eliminates two of the available event outputs.
Putting Innovation to the Test

Copyright © 1996–  TestEquity LLC.