Model 1007S Temperature Chamber Specifications

Temperature Range
-35°C to +175°C (optional to +205°C)
Control Tolerance
±0.5°C (Short-term variations measured at the control sensor after stabilization)
Uniformity
±1°C (Variations throughout the chamber after stabilization)
 
Cool Down Transition Time (empty)*
Start Temp End Temp
+23°C 0°C -10°C -20°C -30°C -35°C
+23°C ---- 1.5 min 3.5 min 7 min 13 min Ultimate
+85°C 5 min 10 min 14 min 18 min 26 min Ultimate
+150°C 12 min 19 min 22 min 26 min 33 min Ultimate
Cool Down Transition Time (with 25 lb. aluminum load)*
Start Temp End Temp
+23°C 0°C -10°C -20°C -30°C -35°C
+85°C 10 18 min 22 min 29 min 41 min Ultimate
+150°C 21 33 min 38 min 43 min 56 min Ultimate
Heat Up Transition Time (empty)*
Start Temp End Temp
+23°C +50°C +85°C +125°C +150°C +175°C
+23°C ----- 1.5 min 7 min 14 min 20 min 25 min
0°C 1.5 min 3.5 min 13 min 20 min 23 min 31 min
-10°C 2.2 min 4.2 min 14 min 22 min 25 min 33 min
-35°C 5.5 min 10 min 16 min 23 min 29 min 34 min

Rate Of Change
To calculate rate of change for a particular condition, take the difference between the Start Temp and End Temp and divide by the Transition Time.
Cool Down Example (empty): From +85°C to -10°C = 95°C / 14 min = 6.8°C/min.
Cool Down Example (with 25 lb. load): From +85°C to -10°C = 95°C / 22 min = 4.3°C/min.
Heat Up Example: From -35°C to +85°C = 120°C / 16 min = 7.5°C/min.

*Note: Transition times are measured after a 2-hour soak at the respective start temperature with an empty chamber, as indicated on the temperature controller, 23°C ambient. Measured with setpoint beyond the start and end temperatures. Does not include the effect of proportional band when approaching setpoint. Performance is reduced by 17% with 50 Hz input power.

 
Live Load Capacity
+23°C 0°C -10°C -20°C -30°C
1,400 Watts 1,000 Watts 850 Watts 625 Watts 300 Watts

 
Refrigeration and Heating System
Refrigerant R-404A (Dupont HP62)
Compressors 1.5 HP Tecumseh hermetic
Condenser Air Cooled
Heat of Rejection 15,200 BTUH (maximum rated chamber load at maximum cooling rate from high temperature soak)
Heater Power 2,000 Watts
Air Flow 450 cfm

 
Instrumentation
Temperature Controller Watlow F4T Touch Screen Controller with RS-232, Ethernet interface, 4.3" color graphic touch screen. OR...
Watlow F4 Controller with RS-232 interface, LED readout of temperature, LCD display of other parameters.
Limit Controller Independent high and low temperature limits. Triggers an audible alarm and shuts down the chamber. Relay contacts provide a safety npower interlock for test sample.
Chart Recorder (Optional) Honeywell DR4300 Series. One pen, 10" circular chart. Mounts in lower front door. Honeywell Chart Recorder

 
Input Power Requirements
230 V ±10%,
60 Hz, 1 PH
Max Current Draw 22 A; Recommended Service 30 A
208 V -5/+10%,
60 Hz, 1 PH
Max Current Draw 22 A; Recommended Service 30 A

Input may be configured for 230 V or 208 V in the field by changing jumpers. Single phase input only. Call for other voltages or 50 Hz operation. Performance is reduced by 17% with 50 Hz input power.

Customer power source must be hard-wired to the chamber by a qualified electrician. Power cord and plug are not included.

Physical Characteristics and Safety
Inside Dimensions 24" W x 21" H x 24" D (7 cubic feet)
609 mm W x 533 mm H x 609 mm D (198 liters)
Outside Dimensions* 33" W x 67.75" H x 54.25" D (nominal)
838 mm W x 1721 mm H x 1378 mm D
Door latch adds 3" to width on right side (may be removed to permit move-in through a 36" doorway). Circulator motor and housing adds 6" to height.
Minimum Installed Clearance 18" from the left and right side
24" from the rear
Window Viewing Area 13.375" W x 9" H
Access Ports 4" Port on left and right side (two total)
Supplied with foam plugs
Weight Chamber Weight: 755 pounds
Shipping Weight: 916 pounds
Sound Level 62 dBA in cooling mode (A-weighted, measured 36" from the front or side surface, 63" from the floor, in a free-standing environment)

 

NOTE: Performance is typical and based on operation at 23°C (73°F) ambient and nominal input voltage. Designed for use in a normal conditioned laboratory. Operation at higher ambient temperatures may result in decreased cooling performance. Additional ports and shelves will also affect performance. Operation above 30°C (85°F) or below 16°C (60°F) ambient is not recommended.