Model 1016S Temperature Chamber Specifications


Temperature Range
-35°C to +175°C
Control Tolerance
±0.5°C (Short-term variations measured at the control sensor after stabilization)
Uniformity
±1.0°C (Variations throughout the chamber after stabilization, up to +8°C)
 
Cool Down Transition Time (empty)*
End Temperature
+23°C 0°C -10°C -20°C -30°C -35°C
Time from +23°C:
°C/minute:
__ 1 min
23°C/min
2 min
16.5°C/min
4 min
10.8°C/min
8 min
6.6°C/min
Ultimate
Time from +85°C:
°C/minute:
4 min
15.5°C/min
7 min
12°C/min
9 min
10.6°C/min
12 min
8.8°C/min
17 min
6.8°C/min
Ultimate
Heat Up Transition Time (empty)*
-30°C to +85 °C 13 minutes (8.8°C/minute)

*Note: Transition times are measured after a 2 hour soak at the respective start temperature with an empty chamber, as indicated on the temperature controller, 23°C ambient. Measured with setpoint beyond the end temperatures with an instantaneous step change in Set Point without a controlled ramp-rate. Does not include the effect of proportional band when approaching setpoint. Performance is reduced by 17% with 50 Hz input power.


 
Live Load Capacity
+23°C 0°C -10°C -20°C -30°C
4,000 Watts 2,800 Watts 2,300 Watts 1,800 Watts 1,550 Watts

 
Refrigeration and Heating System
Refrigerant R-404A (Dupont HP-62)
Compressors 3.5 HP Copeland scroll compressor. More about Scroll Compressors >>
Condenser Air Cooled
Heat of Rejection 30,200 BTUH (maximum rated chamber load at maximum cooling rate from high temperature soak)
Heater Power 4,200 Watts @ 208 V input
Air Flow 830 cfm
Instrumentation
Temperature Controller Watlow F4T Touch Screen Controller with RS-232, Ethernet interface, 4.3" color graphic touch screen. OR...
Watlow F4 Controller with RS-232 interface, LED readout of temperature, LCD display of other parameters.
Limit Controller Independent high and low temperature limits. Triggers an audible alarm and shuts down the chamber. Relay contacts provide a safety power interlock for test sample.
Chart Recorder (Optional) Honeywell DR4300 Series. One pen, 10" circular chart. Mounts in lower front door.

 
Input Power Requirements
230 V ±10%, 60 Hz, 3 Phase Max Current Draw 31 A; Recommended Service 40 A
208 V -5/+10%, 60 Hz, 3 Phase Max Current Draw 26 A; Recommended Service 35 A

Input may be configured for 230 V or 208 V in the field by changing jumpers. Three phase load is balanced. Call for other voltages or 50Hz operation. Performance is reduced by 17% with 50Hz input power.


Customer power source must be hard-wired to the chamber by a qualified electrician. Power cord and plug is not included.


Physical Characteristics and Safety
Inside Dimensions 30" W x 30" H x 30" D (15.6 cubic feet)
762 mm W x 762 mm H x 762 mm D (442 liters)
Outside Dimensions 38" W x 78.5" H x 56" D (nominal)
914 mm W x 1994 mm H x 1422 mm D
Door latch adds 3" to width on right side.
Minimum Installed Clearance 18" from the left and right side
24" from the rear
Window Viewing Area 18" W x 12" H
Access Ports 4" Port on left and right side (two total). Supplied with foam plugs.
Weight Chamber Weight: 1,080 pounds
Shipping Weight: 1,220 pounds
Sound Level 67 dBA in cooling mode (A-weighted, measured 36" from the front surface, 63" from the floor, in a free-standing environment)

NOTE: Performance is typical and based on operation at 23°C (73°F) ambient and nominal input voltage. Designed for use in a normal conditioned laboratory. Operation at higher ambient temperatures may result in decreased cooling performance. Additional ports and shelves will also affect performance. Operation above 30°C (85°F) or below 16°C (60°F) ambient is not recommended.


Due to continuous product improvements, specifications subject to change without notice.