Model 1027S Temperature Chamber Specifications
Cool Down Transition Time (empty)* | ||||||
Start Temp | End Temp | |||||
+23°C | 0°C | -10°C | -20°C | -30°C | -35°C | |
+150°C | 9.0 min | 14 min | 17 min | 20 min | 26 min | Ultimate |
+85°C | 4.0 min | 7.5 min | 10 min | 13 min | 19 min | |
+23°C | ----- | 1 min | 2 min | 4 min | 9 min |
Cool Down Transition Time (with 80 lb. aluminum load)* | ||||||
Start Temp | End Temp | |||||
+23°C | 0°C | -10°C | -20°C | -30°C | -35°C | |
+150°C | 20.0 min | 28 min | 33 min | 38 min | 46 min | Ultimate |
+85°C | 8.0 min | 16 min | 20 min | 25 min | 33 min |
Heat Up Transition Time (empty)* | ||||||
Start Temp | End Temp | |||||
+23°C | +50°C | +85°C | +125°C | +150°C | +175°C | |
+23°C | ----- | 1.5 min | 7 min | 14 min | 20 min | 25 min |
0°C | 1.5 min | 3.5 min | 13 min | 20 min | 23 min | 31 min |
-10°C | 2.2 min | 4.2 min | 14 min | 22 min | 25 min | 33 min |
-35°C | 5.5 min | 10 min | 16 min | 23 min | 29 min | 34 min |
Heat Up Transition Time (with 80 lb. aluminum load)* | ||||||
Start Temp | End Temp | |||||
+23°C | +50°C | +85°C | +125°C | +150°C | +175°C | |
0°C | 5 min | 13 min | 23 min | 36 min | 45 min | 55 min |
-35°C | 11 min | 19 min | 29 min | 42 min | 51 min | 61 min |
Rate Of Change
To calculate rate of change for a particular condition, take the difference between the Start Temp and End Temp and divide by the Transition Time.
Cool Down Example (empty): From +85°C to 0°C = 85°C / 7.5 min = 11.3°C/min.
Cool Down Example (with 80 lb. load): From +85°C to 0°C = 85°C / 16 min = 5.3°C/min.
Heat Up Example (empty): From 0°C to +85°C = 85°C / 13 min = 6.5°C/min.
Heat Up Example (with 80 lb. load): From 0°C to +85°C = 85°C / 23 min = 3.7°C/min.
*Note: Transition times are measured after a 2 hour soak at the respective start temperature with an empty chamber, as indicated on the temperature controller, 23°C ambient. Measured with setpoint beyond the start and end temperatures. Does not include the effect of proportional band when approaching setpoint. Performance is reduced by 17% with 50 Hz input power.
Live Load Capacity
+23°C | 0°C | -10°C | -20°C | -30°C |
4,000 Watts | 2,800 Watts | 2,300 Watts | 1,800 Watts | 1,250 Watts |
Refrigeration and Heating System
Refrigerant | R-404A (Dupont HP-62) | |
Compressors | 3.5 HP Copeland scroll compressor. More about Scroll Compressors >> | |
Condenser | Air Cooled | |
Heat of Rejection | 30,200 BTUH (maximum rated chamber load at maximum cooling rate from high temperature soak) | |
Heater Power | 4,000 Watts @ 208 V input | |
Air Flow | 830 cfm |
Instrumentation
Temperature Controller | Watlow F4T Touch Screen Controller with RS-232, Ethernet interface, 4.3" color graphic touch screen. OR... Watlow F4 Controller with RS-232 interface, LED readout of temperature, LCD display of other parameters. |
|
Limit Controller | Independent high and low temperature limits. Triggers an audible alarm and shuts down the chamber. Relay contacts provide a safety power interlock for test sample. | |
Chart Recorder | (Optional) Honeywell DR4300 Series. One pen, 10" circular chart. Mounts in lower front door. |
Input Power Requirements
230 V ±10%, 60 Hz, 3 Phase | Max Current Draw 31 A; Recommended Service 40 A | |
208 V -5/+10%, 60 Hz, 3 Phase | Max Current Draw 26 A; Recommended Service 35 A | |
Input may be configured for 230 V or 208 V in the field by changing jumpers. Three phase load is balanced. Call for other voltages or 50 Hz operation. Customer power source must be hard-wired to the chamber by a qualified electrician. Power cord and plug is not included. |
Physical Characteristics and Safety
Inside Dimensions | 40" W x 32" H x 36.5" D (27 cubic feet) 1016 mm W x 812.8 mm H x 927 mm D (764 liters) |
|
Outside Dimensions | 49" W x 73.25" H x 63" D (nominal) 1244 mm W x 1860 mm H x 1600 mm D Door latch adds 3" to width on right side. Circulator motor and housing adds 6" to height - may be removed for move-in. |
|
Minimum Installed Clearance | 18" from the left and right side 24" from the rear |
|
Window Viewing Area | 16" W x 13" H | |
Access Ports | 4" Port on left and right side (two total) Supplied with foam plugs |
|
Weight | Chamber Weight: 1,410 pounds Shipping Weight: 1,595 pounds |
NOTE: Performance is typical and based on operation at 23°C (73°F) ambient and nominal input voltage. Designed for use in a normal conditioned laboratory. Operation at higher ambient temperatures may result in decreased cooling performance. Additional ports and shelves will also affect performance. Operation above 30°C (85°F) or below 16°C (60°F) ambient is not recommended. Performance is reduced by 17% with 50 Hz input power.
Due to continuous product improvements, specifications subject to change without notice.